The Thermal Effects of Kirschner Wire Fixation on Small Bones

Achal Khanna, Sotirios Plessas, Paul Barrett, L.C. Bainbridge

Journal of Hand Surgery (European Volume) · 1999 · 34 citations · 10 references

Concepts

Abstract

The aim of this study was to assess bone heating caused by the passage of fine (<2 mm) K-wires of different types. Stainless steel K-wires of trocar and diamond point configurations (0.8-2.0 mm) were drilled into the metatarsal bones from a freshly amputated lower limb at a constant force. Temperature measurements were made by miniature thermocouples inserted into the bones, at 3 second intervals over a period of 3 minutes while each K-wire was drilled three times. The temperature reached varied with the tip configurations and the diameter of the K-wires. Regardless of point configuration, thinner wires generated more heat than thicker ones.

References

10