Physics in Medicine and Biology · 2008 · 74 citations · 21 references
MedicineRadiation OncologyRadiation BiophysicsRadiation DosimetryThermal AnalysisPatient-specific DosimetryRadiation Dose MeasurementRadiation TherapyThree-dimensional Dose DistributionBody TemperatureExternal Beam Radiotherapy DosimetryFormaldehydeMeltingMelting PointMelting Point AnalysisPhysical GelsThree-dimensional Dose Calculation
Polymeric gel dosimeters are being used to verify three-dimensional (3D) dose distributions of different types of radiotherapy treatments, especially the most complexes ones. An important factor that can limit the wider use of this kind of dosimeter is temperature, as gel melting can destroy 3D information. This work shows that adding formaldehyde to the gel preparation increases the melting point, allowing its use in warmer environments, including up to body temperature. An addition of 3% in mass of the formaldehyde solution to a MAGIC type gel dosimeter increased its melting point from 25 to 69 degrees C. Also important were a 12.5% increase in gel sensitivity and an expressive decrease in relaxation rate R2 uncertainty.
21
A basic study of some normoxic polymer gel dosimeters
Yves De Deene, Christopher A. Hurley, A Venning et al. · Physics in Medicine and Biology · 2002 · 275 citations
Yves De Deene, K Vergote, Carolien Claeys et al. · Physics in Medicine and Biology · 2006 · 274 citations
Investigation of the PAGAT polymer gel dosimeter using magnetic resonance imaging
A Venning, Brendan Hill, S Brindha et al. · Physics in Medicine and Biology · 2005 · 204 citations