Publication | Closed Access
Rapid Determination of Sildenafil Citrate in Pharmaceutical Preparations Using Monolithic Silica HPLC Column
33
Citations
15
References
2003
Year
High Throughput AnalysisMedicinal ChemistryPharmaceutical ScienceTherapeutic Drug MonitoringRapid DeterminationIsocratic ElutionMedicineDrug PurityBioanalysisSildenafil CitratePharmaceutical TechnologyAnalytical ChemistryLiquid ChromatographyClinical ChemistryChromatographic AnalysisPharmacologyChromatographyDrug Analysis
Abstract A simple, rapid, and sensitive high performance liquid chromatography (HPLC) method has been developed and validated for the determination of sildenafil citrate in pharmaceutical preparations. The method was developed utilizing a 100 × 4.6 mm I.D. monolithic silica column and an isocratic elution of 60:40, v/v acetonitrile/water. The elution of the analyte was monitored at 292 nm with a flow rate of 2.0 mL/min. All analyses were conducted at ambient temperature. Linearity was observed in the concentration range from 50–3000 ng/mL, with a correlation coefficient (R 2) greater than 0.999. The limit of detection was 25 ng/mL. Parameters of validation prove the precision of the method and its applicability for the determination of sildenafil citrate in pharmaceutical tablet formulations. The method is fast (less than 1 min) and is suitable for high throughput analysis of the drug.
| Year | Citations | |
|---|---|---|
Page 1