An Internal Fixator for Posterior Application to Short Segments of the Thoracic, Lumbar, or Lumbosacral Spine Design and Testing

Martin H. Krag, Bruce D. Beynnon, Malcolm H. Pope, John W. Frymoyer, Larry D. Haugh, Donald L. Weaver

Clinical Orthopaedics and Related Research · 1986 · 286 citations · 0 references

Concepts

Abstract

A new spinal implant has been designed and biomechanical testing completed, intended for application to "short-segment" spinal defects such as disc degeneration, fracture, spondylolisthesis, or tumor. Major improvements over currently available devices include: only 2-3 vertebrae are spanned, not 5-7 as with Harrington rods; true three-dimensional fixation is achieved, preventing such problems as hook or rod dislocation; three-dimensional adjustment is easily accomplished, allowing fracture or spondylolisthesis reduction to be readily performed; attachment to vertebrae is by means of transpedicular screws eliminating deliberate encroachment into the spinal canal, such as Luque wires or Harrington hooks; no special alignment between screws is needed (such as with holes or slots in a plate), allowing screw placement to fully conform to anatomic structures; and laminectomy sites and lumbosacral junction are readily instrumented. Background investigations presented here for design of this device include: CT-defined pedicle morphometry showing that screws may be larger than those currently used; effect of pitch, minor diameter, and tooth profile on screw pull-out strength; mechanical testing of a compact, three-dimensionally adjustable, strong, nonloosening articulating clamp; and establishing of the relationship between depth of penetration and strength of fixation of transpeduncular screws.