Journal of Chemistry · 2008 · 11 citations · 3 references
PharmacyPharmaceutical AnalysisPharmaceuticsBiopharmaceuticsChromatographic AnalysisPharmaceutical FormulationBioanalytical Method ValidationChromatographic Analysis Of PharmaceuticalsChromatographyLiquid ChromatographyAccuracyHigh-performance Liquid Chromatography Method DevelopmentMobile Phase SelectivityRobustnessRobustness ValidationSensitivitySystem Suitability CriteriaSystem Suitability Testing
A rapid and sensitive RP‐HPLC method with UV detection (262 nm) for routine analysis of adefovir dipivoxil in bulk and in pharmaceutical formulation was developed. Chromatography was performed with mobile phase containing a mixture of acetonitrile and phosphate buffer (50:50, v/v) with flow rate 1.0 mL min -l . In the range of 5.0‐100 µg/mL, the linearity of adefovir dipivoxil shows a correlation co‐efficient of 0.9999. The proposed method was validated by determining sensitivity accuracy, precision, robustness stability, specificity, selectivity and system suitability parameters.
3
Martindale: the complete drug reference
Choice Reviews Online · 2005 · 2.7K citations