Gold Nanoparticle-Modified Indium Tin Oxide Electrode for Highly Sensitive Electrochemical Detection of Melamine

Hosneara Akter, Aasiya Shaikh, T. R. Chowdhury, Mohammad Shafiqur Rahman, P. K. Bakshi, A. J. Saleh Ahammad

ECS Electrochemistry Letters · 2013 · 17 citations · 1 references

Concepts

Abstract

A simple and highly sensitive electrochemical method, based on a gold nanoparticle (GNP)-modified indium tin oxide (ITO) electrode (ITO/GNP), was developed for the detection of melamine (MLM). The modified electrode surface was characterized by scanning electron microscope (SEM) and cyclic voltammetry (CV). MLM was detected based on the adsorption of MLM onto ITO/GNP electrode causing suppression in the cathodic peak current of ferricyanide. A wide linear concentration range from 5 to 500 nM with a lower detection limit of 1.0 nM (i.e. 0.13 ppb) for MLM was obtained. The proposed method was successfully applied to detect MLM in milk samples.

References

1