IEEE Transactions on Biomedical Engineering · 1981 · 18 citations · 15 references
Life SciencesPhysiologyExercise PhysiologyPulmonary Function TestingVentilationGas ExchangeGas Exchange AssessmentGas Exchange PhysiologyLung VolumesPulmonary PhysiologyRespiratory PhysiologyError SensitivityExercise Gas Exchange TestingExpiratory Flow Rate MeasurementFlow-volume RelationshipsFlowmeter CalibrationInspiratory MechanicsLung Volume QuantificationSensitivity
During expiration, fluctuations in gas composition, water vapor, and temperature result in flowmeter errors when the flowmeter is calibrated for a given ambient inspiratory gas. In this paper, we indicate that alternative analytical methods of caleulating breath-to-breath gas exchange exhibit differing sensitivities to this error. A theoretical sensitivity analysis is verified by O2 consumption records from rest to exhausting exercise. We conclude that an error sensitivity of less than one is achieved by a method that incorporates measurements of nitrogen flow into and out of the lung, and includes the analysis of breath-to-breath changes in lung volume.
15
Anaerobic threshold and maximal aerobic power for three modes of exercise
J. A. Davis, Paul Vodak, Jack H. Wilmore et al. · Journal of Applied Physiology · 1976 · 469 citations