Abstract

The new generation of synthetic diamonds grown as a CVD single crystal on a high pressure high temperature substrate offers a wide range of applications. In particular, because of the near tissue equivalence and its small size (good spatial resolution), CVD single crystal diamond finds applicability in radiotherapy as a dosemeter of ionizing radiation. In this paper we report the electrical and dosimetric properties of a new diamond detector which was fabricated at IFJ based on a single crystal detector-grade CVD diamond provided with a novel contact metallization. Diamond properties were assessed at IFJ using a Theratron 680E therapeutic 60 Co gamma rays unit and at COOK with 6 and 18 MV x-rays Varian Clinac CL2300 C/D accelerator. The new dosemeter showed high electric and dosimetric performances: low value of dark current, high current at the level of some nanoamperes during irradiation, very fast dynamic response with a rise time amounting to parts of a second, good stability and repeatability of the current and linearity of the detector signal at different dose and dose rate levels typically applied in radiotherapy. The results confirm the potential applicability of diamond material as a dosemeter for applications in radiotherapy.