Review of Scientific Instruments · 1991 · 87 citations · 6 references
Natural SciencesMaterials SciencePhysicsHeat TransferMetrologyThermodynamicsCalorimetryHeat Capacity MeasurementHeat ConductionTemperature MeasurementTemperature MetrologyThermal Properties Of MaterialsEffective Thermal ConductivityHeat CapacityLow-temperature Heat CapacityMeasurementNoncontact MeasurementProcess ParametersThermal ConductivityThermal Conductivity Measurements
A concept is developed and described which allows to measure the heat capacity and the effective thermal conductivity of stable and undercooled liquid metals and alloys in an electromagnetic levitation apparatus. We propose to use an ac pulse heating method which is used nowadays as a standard technique for precision measurement of low temperature heat capacities. The ideal process parameters including the drop diameter D, temperature T, and frequency of measurement ω can be optimized when the following relations hold for the external and internal relaxation time constants τ1 and τ2, respectively: ωτ1≳10 and ωτ2<0.1. Then heat capacity data can be obtained with an accuracy of better than 1% with D about 5 to 10 mm, T between 1200 and 1800 K and ω between 0.1 and 1 Hz for typical metals and alloys.
6
Steady-State, ac-Temperature Calorimetry
Paul F. Sullivan, G. M. Seidel · Physical Review · 1968 · 750 citations
Heat Capacity Measurements on Small Samples at Low Temperatures
Rainer Bachmann, F. J. DiSalvo, T. H. Geballe et al. · Review of Scientific Instruments · 1972 · 553 citations
慶一 石原 · Medical Entomology and Zoology · 1987 · 180 citations