Separating the Laparoscopic Camera Cord From the Monopolar “Bovie” Cord Reduces Unintended Thermal Injury From Antenna Coupling

14

Citations

10

References

2015

Year

Abstract

Separation of the laparoscopic camera cord from the active electrode cord decreases thermal injury from antenna coupling at the camera trocar incision in comparison to the parallel orientation of these cords. Therefore, parallel orientation of these cords (an arrangement promoted by integrated operating rooms) should be abandoned. The findings of this study should influence the operating room setup for all laparoscopic cases.