Publication | Closed Access
Separating the Laparoscopic Camera Cord From the Monopolar “Bovie” Cord Reduces Unintended Thermal Injury From Antenna Coupling
14
Citations
10
References
2015
Year
Separation of the laparoscopic camera cord from the active electrode cord decreases thermal injury from antenna coupling at the camera trocar incision in comparison to the parallel orientation of these cords. Therefore, parallel orientation of these cords (an arrangement promoted by integrated operating rooms) should be abandoned. The findings of this study should influence the operating room setup for all laparoscopic cases.