Journal of Cataract & Refractive Surgery · 1988 · 10 citations · 5 references
MedicineOphthalmologyOptometryOphthalmic SurgeryOpticsVisual OpticsCataract AssessmentCataract EvaluationCorneal OpticsOcular BiometryRefractive ErrorRefractive Error AssessmentIol Power CalculationLens BiometryLens Power Determination
A method to measure the power of an intraocular lens (IOL) in situ using an extended range keratometer and a depth measuring apparatus is described. With this method, the IOL power was measured in ten patients with an anterior chamber IOL and in ten patients with a posterior chamber IOL and compared with the labeled power. The total mean error (measured power minus labeled power) was found to be 0.01 diopters (D) with a standard deviation of 0.22 D. The largest error was 0.40 D.
5
On the calculation of power from curvature of the cornea.
Thomas Olsen · British Journal of Ophthalmology · 1986 · 249 citations · Full text
Measurement of Corneal Thickness With the Haag-Streit Pachometer
S Mishima, Bengt O. Hedbys · Archives of Ophthalmology · 1968 · 175 citations
Medicine, Ophthalmology, Optometry +10