Publication | Closed Access
Physiologic Monitoring Alarm Load on Medical/Surgical Floors of a Community Hospital
58
Citations
8
References
2011
Year
1) Standard critical care alarm limits appear be too sensitive for subacute care areas of the hospital. 2) For most patients these alarm limits do not create a significant alarm load; however, for a small number of patients they cause a significant alarm load. 3) Alarm loads can be controlled with alarm limit settings appropriate to the population. 4) Current technology for HR and SpO(2) appear suitable for continuous monitoring of this population.