PubMed · 1988 · 124 citations · 6 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryBone GraftingJoint Replacement SurgeryOrthopedicsCement-retained ProsthesesSocket DesignBone Cement FixationCemented FixationRevision Arthroplasty CementationSocket Stability
In a series of 200 Charnley low friction arthroplasties revised for various reasons, the preoperative anteroposterior roentgenograms were categorized in terms of demarcation at the bone-cement interface of the socket and compared with the operative findings of movement at the bone-cement junction. All sockets showing no demarcation were found to be fixed soundly. Seven percent of the sockets showing demarcation of the outer one-third only were loose, and when two-thirds of the sockets were demarcated, 71% were loose. Ninety-four percent of the sockets with complete demarcation and all sockets that had migrated were found to be loose at revision. There is a definite correlation between roentgenographic demarcation and socket loosening. The more extensive the demarcation, the more likely it is that the socket is loose. Radiographic demarcation of the cemented socket is a prognostic sign for eventual failure.
6
Robert D. Beckenbaugh, D M Ilstrup · Journal of Bone and Joint Surgery · 1978 · 362 citations
Eduardo García-Cimbrelo, Vicente Diez-Vazquez, Rosário Madero et al. · Journal of Bone and Joint Surgery · 1997 · 72 citations