Arthroscopy The Journal of Arthroscopic and Related Surgery · 1998 · 16 citations · 11 references
MedicineOrthopedic SurgeryInstrumentationFixation BiomechanicsInternal FixationJoint StrengthInterference Screw MaterialsScrew FixationScrew Fixation MechanicsScrew Pullout Strength
Small-diameter interference screws are a widely used method of securing bone-tendon grafts during anterior cruciate ligament reconstructive procedures. Recent reports document tunnel widening and subsequent loss of fixation, leading to a need for alternative fixation methods. In cadaver tissue, the fixation strength of large-diameter (11-, 13-, and 15-mm) interference screws was tested and compared with that of clinically used 9-mm diameter screws. Patellar-tendon sections were fashioned into three grafts, each with bone plugs measuring 10 x 20 x 7.5 mm. Tunnels were created in the epiphyseal bone of the femurs and tibias, resulting in gap sizes of 4.5-, 6.5-, 8.5-, and 10.5-mm for the 9-, 11-, 13-, and 15-mm diameter screws, respectively. Grafts were pulled from the tunnel at a rate of 51 cm/min. Average fixation strengths were 276, 302, 328, and 328 N for the 9-, 11-, 13-, and 15-mm screws, respectively. No significant difference existed between the fixation strengths (ANOVA, P < or = .05).
11
Daniel D. Buss, Russell F. Warren, T L Wickiewicz et al. · Journal of Bone and Joint Surgery · 1993 · 258 citations