Publication | Closed Access
Reproducibility of Measurements With the Nerve Fiber Analyzer (NFA/GDx)
66
Citations
0
References
2000
Year
The reproducibility of measurements varied across parameters. In general, the reproducibility of measurements with the Nerve Fiber Analyzer was high. The reproducibility of measurements was similar between healthy patients and patients with glaucoma. Any measured change in nerve fiber layer thickness would be statistically significant if it exceeded approximately 7 or 8 microm in the superior maximum or inferior maximum parameter in healthy patients. Reproducibility of measurements hardly differed between single images and mean images. The reproducibility of measurements among the three instruments we used was highest for straight parameters.