Otology & Neurotology · 2013 · 56 citations · 12 references
MedicineOtolaryngologySurgeryClinical AudiologyCochlear ImplantsOtologyAdult Cochlear ImplantsCochlear ImplantationHearing LossPediatric Cochlear ImplantationPediatric Cochlear ImplantsSurgical ImplantationCochlear Implant Revision SurgeryCochlear Implantation In ChildrenCochlear Implants In ChildrenOperative MorbidityReimplantation SurgeryRevision Cochlear Implant SurgeryRevision Cochlear ImplantationRevision Surgery Rates
OBJECTIVE: To evaluate the long-term adverse cochlear implant (CI) outcomes resulting in revision surgery including CI reimplantation (CIR). PATIENTS: Pediatric and adult patients requiring revision procedures after CI placement. INTERVENTION(S): Revision surgery on cochlear implant patients. MAIN OUTCOME MEASURES: Device type, length of total device follow-up, time to device failure, cause for failure, peak pre-CIR and post-CIR audiometric performance, rate of surgical site complications, and operative findings. RESULTS: A total of 317 patients, receiving 439 CIs between January 2000 and April 2012, met inclusion criteria for this series. For the patients implanted at our institution, the revision surgery rate was 4.1%, with a CIR rate of 3.0%. The CIR rates among the pediatric and adult populations were 5.0% and 1.3%, respectively (p = 0.0336). The rate of revision procedures because of failed fixation or device extrusion was 0.9%. Device failure was experienced in 8 patients in our series, with 75% occurring with the CI24R (CS) device. CONCLUSION: All reimplanted patients with available data had good audiometric outcomes, with the exception of those reimplanted for soft failure who had poor immediate auditory function. Using the manufacturers' recommended surgical technique, including drilling a bony recess with suture fixation, very low surgical revision rates were achieved. Pediatric patients experienced significantly higher complications requiring CIR. All hard failures in this series occurred in the pediatric group and in a single device. Continued follow-up will be needed to determine if additional devices will succumb to this mode of failure.
12
Complications of cochlear implant surgery.
Ronald A. Hoffman, Nancy L. Cohen · PubMed · 1995 · 258 citations
Medicine, Otolaryngology, Surgery +12
Cochlear Implant Failures and Revision
Amy Anne D. Lassig, Teresa A. Zwolan, Steven A. Telian · Otology & Neurotology · 2005 · 194 citations
Medicine, Otolaryngology, Surgery +12
Incidence and indications for revision cochlear implant surgery in adults and children
Kevin D. Brown, Sarah S. Connell, Thomas J. Bałkany et al. · The Laryngoscope · 2008 · 152 citations
Medicine, Otolaryngology, Surgery +17
Cochlear Reimplantation: Causes of Failure, Outcomes, and Audiologic Performance
Mathieu Côté, P. Ferron, François Bergeron et al. · The Laryngoscope · 2007 · 148 citations
Revision Cochlear Implant Surgery in Children
Robert D. Cullen, José N. Fayad, William M. Luxford et al. · Otology & Neurotology · 2008 · 135 citations
Medicine, Otolaryngology, Surgery +17