<title>Application of long-period grating sensors to respiratory function monitoring</title>

T. Allsop, Tim Earthrowl, Richard Revees, D. J. Webb, Martin R. Miller, Barrie W. Jones, I. Bennion

Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · 2004 · 12 citations · 17 references

Concepts

Abstract

A series of in-line curvature sensors on a garment are used to monitor the thoracic and abdominal movements of a human during respiration. These results are used to obtain volumetric tidal changes of the human torso showing reasonable agreement with a spirometer used simultaneously to record the volume at the mouth during breathing. The curvature sensors are based upon long period gratings written in a progressive three layered fibre that are insensitive to refractive index changes. The sensor platform consists of the long period grating laid upon a carbon fibre ribbon, which is encapsulated in a low temperature curing silicone rubber. An array of sensors is also used to reconstruct the shape changes of a resuscitation manikin during simulated respiration. The data for reconstruction is obtained by two methods of multiplexing and interrogation: firstly using the transmission spectral profile of the LPG's attenuation bands measured using an optical spectrum analyser; secondly using a derivative spectroscopy technique.

References

17