Dose Reduction and Image Quality Assessment in 64-Detector Row Computed Tomography of the Coronary Arteries Using an Automatic Exposure Control System

Marco Francone, Elisabetta Di Castro, Alessandro Napoli, Chiara Bolzan, Iacopo Carbone, Linda Bertoletti, Luigi Iuliano, Carlo Catalano, Roberto Passariello

Journal of Computer Assisted Tomography · 2008 · 10 citations · 33 references

Concepts

Abstract

OBJECTIVE: To evaluate dose reduction and image quality in coronary 64-slice multidetector computed tomography using an automatic exposure control system (AECs). METHODS: A total of 101 patients were divided into 4 groups. Tube current was 600 and 800 mAs in groups A and B and adapted at 600 and 800 quality-reference mAs using an AECs in groups C and D. Effective dose and organ-equivalent dose were evaluated. Image noise was quantified as standard deviation of air-space attenuation. Two observers assessed technical adequacy and image quality using a 4-point scale. RESULTS: Effective dose ranged from 8.6 mSv (group C) to 15 mSv (group B) with significant dose reduction for examinations performed at 600 mAs (21.7%) and 800 mAs (29.4%). Contribution of organ-equivalent doses showed higher exposure for lungs (42%) and breast (22%). Noise was significantly higher in groups studied with AECs. Larger coronary segments resulted in higher image quality scores without differences between groups. CONCLUSION: Automatic exposure control systems provides images of diagnostic quality with substantial dose reduction.