Electrophoresis · 2004 · 69 citations · 27 references
Natural SciencesChemical EngineeringChemistryAnalytical ChemistryElectroanalytical MethodsElectrochemistryElectroanalytical ChemistryElectrochemical DetectionElectrolyte AnalysisElectrolyte SolutionsIon-selective ElectrodesAmino AcidsAnalytical Detection LimitsAnalytical ElectrophoresisElectrical Conductivity MeasurementsElectrochemical SensingElectrophoresisElectrophoretic DetectionMicrochip Electrophoresis
The detection of underivatized anionic sulfonates, carboxylates, amino acids, sugars, and artificial sweeteners, and of cationic dopamine, ephedrine, and metanephrine in microfabricated electrophoresis devices is demonstrated. This was achieved by high-voltage contactless conductivity measurements with external electrodes. Poly(methyl methacrylate) chips with thin covers to enable sensitive contactless detection were used for most determinations but glass microchips had to be employed for amino acids and sugars. The plastic chips were found not be stable in the alkaline media required to render those two classes of species in the ionic form amenable for separation and detection. The reproducibility of peak area measurements was about 1% or better and the detection limits ranged between 1 and 30 microM for the different compounds examined.
27
Contactless Conductivity Detection for Capillary Electrophoresis
Andreas Zemann, E. Schnell, Dietmar Volgger et al. · Analytical Chemistry · 1998 · 538 citations
Contactless Conductivity Detector for Microchip Capillary Electrophoresis
Martin Pumera, Joseph Wang, František Opekar et al. · Analytical Chemistry · 2002 · 230 citations