ECS Transactions · 2013 · 12 citations · 0 references
EngineeringMaterials EngineeringNuclear EngineeringAnalytical ChemistryNuclear InstrumentationNuclear MaterialsAlloysDetector SystemsMass AnalyzersMass SpectrometryMetalsNuclear Fuel MaterialsElectron HeatingMagnetic Sector AnalyzersMagnetic Sector DetectorsMetallic FuelNuclear Fuel FabricationQuadrupole Mass AnalyzersQuadrupole Mass SpectrometryVaporization
Benefiting hugely from many years of experience of conducting vaporization studies with a commercially bought magnetic sector instrument (VG Micromass 30 BK), we designed and constructed a KEMS system having a quadrupole mass spectrometer. It essentially consisted of two chambers, one housing the Knudsen cell furnace assembly and the other housing the QMS with an indigenously designed and fabricated vacuum-isolation valve between these two chambers. The QMS initially used was from Extrel, but later that from Hiden was installed. Systematic vaporization experiments were undertaken on iron, nickel, chromium, zirconium, a Fe-Al alloy, MnO, uranium, and a U-Zr alloy. Electron bombardment heating was employed and the temperatures were measured by sighting the optical pyrometer onto a blackbody hole drilled at the bottom of the Knudsen cell. The results on the elements and the Fe-Al alloy were satisfactory while those on MnO, U, and U-Zr alloy were disconcerting but interesting and educative.