Publication | Open Access
Metal-on-metal bearing in hip prosthesis generates 100-fold less wear debris than metal-on-polyethylene
157
Citations
18
References
1999
Year
Aseptic loosening due to osteolysis in total hip replacement has been related to wear debris released from prosthetic components. Retrospective longterm observations of patients with the metal-on-metal prosthesis has shown long-term survivorship and good mechanical performance. Thus, the new and modified metal-on-metal prosthesis has been introduced on the market. Historical clinical data from the 1st generation metal-on-metal hip prosthesis may not be relevant for the 2nd generation of metal-on-metal hip prosthesis. Therefore, preclinical testing of the prosthesis must be conducted before clinical evaluation. We assessed the tribological performance of the metal-on-metal prosthesis versus the metal-on-polyethylene prosthesis introduced on the market as Metasul and Protasul, respectively. In a 12-channel joint simulator, 6 metal-on-metal bearing and 3 metal on polyethylene prostheses were tested, with the same number of corresponding soak controls. The wear was assessed gravimetrically. The "steady-state" wear-rates from the metal-on-metal prosthesis were almost 100 times less than that from the metal-on-polyethylene prosthesis. The tribological wear performance of the metal-on-metal hip prosthetic system is promising.
| Year | Citations | |
|---|---|---|
Page 1