Neurosurgery · 2004 · 10 citations · 15 references
MedicineSurgeryVeterinary SurgeryCerebrovascular SurgerySmall Animal SurgeryVascular SurgeryBalloon AngioplastyCarotid Artery StentingCarotid Artery SurgeryEndovascular SurgeryVascular TraumaBalloon-expandable StentsCommon Carotid ArteryEndovascular Vascular RepairNeurovascular Device EvaluationNeurovascular SafetyVascular Access DevicesVascular SpasmVascular Stenting
OBJECTIVE: We developed an animal model to evaluate vascular trauma induced by endovascular devices that are proposed for the treatment of cerebrovascular diseases. METHODS: The model was constructed in six domestic swine by elongating the common carotid artery using interpositional grafts. Balloon catheters (n = 3), stents (n = 2), and a snare were tested 9 to 13 days after surgery. Device performance was evaluated by angiography, macroscopic photography, and histopathological examination. RESULTS: Animals tolerated the surgical procedure well, and artificial siphons were thought to provide realistic conditions for device testing. Balloon catheters induced minimal trauma, whereas coronary stents caused severe spasm or thrombosis and extensive macroscopic changes. CONCLUSION: Construction of an in vivo siphon model is feasible and potentially useful for testing neurovascular devices.
15
Balloon-expandable intracoronary stents in the adult dog.
Richard A. Schatz, J C Palmaz, Fermin O. Tio et al. · Circulation · 1987 · 375 citations · Full text
Medicine, Cardiology, Surgery +12
Tarik F. Massoud, Cheng Ji, Fernando Viñuela et al. · PubMed · 1994 · 86 citations · Full text
A human cadaveric circulation model
H. Edward Garrett · Journal of Vascular Surgery · 2001 · 86 citations · Full text