PubMed · 1987 · 32 citations · 9 references
PharmacyPharmaceuticsPharmacologyAnalytical ChemistryChromatographic AnalysisMass Spectrometric AnalysisChromatographyColumn TechnologyFluorescence SpectroscopyLiquid ChromatographyMetabolite IdentificationPhotochemistryBioanalytical ChromatographyColumn ChromatographyHigh-performance Liquid Chromatography Method DevelopmentReactive Metabolite Detection
On irradiation with short-wavelength UV light in the presence of hydrogen peroxide, methotrexate and its metabolites 7-hydroxymethotrexate and 2,4-diamino-N10-methylpteroic acid are cleaved into highly fluorescent products. This reaction can be used for the sensitive and selective detection of the compounds in biological fluids, following reversed-phase high-performance liquid chromatographic separation. Study of the effect of the mobile phase composition and irradiation time on fluorescence signal intensity showed that a residence time of ca. 3 s in the on-line photochemical reactor was best. The detection limit for methotrexate was 0.4 ng, for 7-hydroxymethotrexate 1.0 ng and for 2,4-diamino-N10-methylpteroic acid 0.6 ng. The addition of dimethylformamide to the mobile phase enhanced the selectivity of separation.
9
Fluorescence Detection in High Performance Liquid Chromatography
H. Lingeman, W.J.M. Underberg, Akira Takadate et al. · Journal of Liquid Chromatography · 1985 · 102 citations