EngineeringCardiologyElectrical EngineeringElectromagneticsRadiation ProtectionRadiation SafetyElectromagnetic CompatibilityElectromagnetic PowerMedical Radiation ShieldingCutoff FormationElectromagnetic InterferenceElectromagnetic Interference ManagementElectromagnetic Interference MitigationPulse CompatibilityPulse Generator CircuitsPulse Generators
Electromagnetic radiation (EMR) in the environment is a source of potential danger to a patient dependent on a cardiac pulse generator. We exposed a variety of demand pulse generators to a wide range of EMR. Reversion to "fixed rate" and occasionally complete cutoff was observed in several devices at power densities approximating those reported in one large metropolitan area. Practical techniques exist for producing pulse generators that are compatible with today's EMR environment, and steps must be taken by pulse generator manufacturers to reduce the sensitivity of their units.