Fully implantable Otologics MET Carina<sup>™</sup>device for the treatment of sensorineural and mixed hearing loss: Audio-otological results
57
Citations
18
References
2010
Year
We did not record any surgical relevant complication in any of the patients, nor any significant postoperative variation in hearing thresholds, for air conduction or bone conduction, indicating the absence of surgical damage to the cochlea. All the patients demonstrated improvements in speech perception abilities with the device functioning and reported subjective benefits. With regard to the postoperative adverse effects, we had problems with feedback noise, which resolved with minor fitting adjustments in seven cases, while it required a second surgery to change the microphone position in the other patient. In one case a minimal extrusion of the microphone cable occurred requiring a revision surgery, a device failure occurred in one case, requiring substitution, and one patient decided on explantation of the device owing to psychological problems.