Physics in Medicine and Biology · 2014 · 79 citations · 41 references
MedicineBiomedical EngineeringNeurosurgeryFunctional NeurosurgeryUltrasonographyUltrasoundAcoustic SimulationDiagnostic UltrasonographyInterventional UltrasoundTransducer TechnologyUltrasonicsUltrasound PhysicsElastic Wave EquationsEngineering SimulationsFinite Difference ModelingPhased Array TransducersTherapeutic Ultrasound TransducersUltrasound TherapyUltrasound-guided InterventionsWave Equations
Accurate simulation of focused ultrasound neurosurgery depends on precise tissue property data, particularly for skull bone. A grid‑based finite‑difference model coupling acoustic and visco‑elastic wave equations was used to simulate eight sonications in five patients, driving a phased‑array transducer with clinical parameters and comparing predicted focal temperatures and thermal‑focus sizes to measured data. Simulations underestimated focal temperatures by about 24 % and thermal‑focus sizes by 22–40 %, with target offsets of 0.3 mm and 1.6 mm, yet they reliably predicted spot locations and temperature elevations sufficiently for preliminary treatment planning.
A computational model utilizing grid and finite difference methods were developed to simulate focused ultrasound functional neurosurgery interventions. The model couples the propagation of ultrasound in fluids (soft tissues) and solids (skull) with acoustic and visco-elastic wave equations. The computational model was applied to simulate clinical focused ultrasound functional neurosurgery treatments performed in patients suffering from therapy resistant chronic neuropathic pain. Datasets of five patients were used to derive the treatment geometry. Eight sonications performed in the treatments were then simulated with the developed model. Computations were performed by driving the simulated phased array ultrasound transducer with the acoustic parameters used in the treatments. Resulting focal temperatures and size of the thermal foci were compared quantitatively, in addition to qualitative inspection of the simulated pressure and temperature fields. This study found that the computational model and the simulation parameters predicted an average of 24 ± 13% lower focal temperature elevations than observed in the treatments. The size of the simulated thermal focus was found to be 40 ± 13% smaller in the anterior-posterior direction and 22 ± 14% smaller in the inferior-superior direction than in the treatments. The location of the simulated thermal focus was off from the prescribed target by 0.3 ± 0.1 mm, while the peak focal temperature elevation observed in the measurements was off by 1.6 ± 0.6 mm. Although the results of the simulations suggest that there could be some inaccuracies in either the tissue parameters used, or in the simulation methods, the simulations were able to predict the focal spot locations and temperature elevations adequately for initial treatment planning performed to assess, for example, the feasibility of sonication. The accuracy of the simulations could be improved if more precise ultrasound tissue properties (especially of the skull bone) could be obtained.
41
A Pilot Study of Focused Ultrasound Thalamotomy for Essential Tremor
W. Jeffrey Elias, Diane Huss, Tiffini Voss et al. · New England Journal of Medicine · 2013 · 865 citations · Full text
MR-guided focused ultrasound thalamotomy for essential tremor: a proof-of-concept study
Nir Lipsman, Michael L. Schwartz, Yuexi Huang et al. · The Lancet Neurology · 2013 · 604 citations