Microfluidic chip for plasma separation from undiluted human whole blood samples using low voltage contactless dielectrophoresis and capillary force

Chia-Chern Chen, Po-Hsiu Lin, C.K. Chung

Lab on a Chip · 2014 · 38 citations · 18 references

Concepts

Abstract

A plasma separating biochip is demonstrated using a capillary-driven contactless dielectrophoresis method with low voltage (~1 V) and high frequency induced electrostatics between red blood cells. The polarized red blood cells were aggregated and separated from plasma with a 69.8% volume separation and an 89.4% removal rate of red blood cells.

References

18