PharmacyBiochemistryChemistryAnalytical ChemistryElemental AnalysisMass Spectrometric AnalysisDetector SystemsInstrumental AnalysisMass AnalyzersMass Spectrometric Elemental AnalysisMass SpectrometryStable Isotope AnalysisIsotope Dilution AnalysisIsotope Dilution Mass SpectrometryIsotope Ratio AnalysisIsotope Ratio Mass SpectrometryMass Spectrometric Isotope AnalysisNitrogen Isotope AnalysisStable Isotope Dilution AnalysisThroughput
A sequential sample treatment system is described that permits use of an isotope-ratio mass spectrometer for automated Rittenberg analyses of ammonium salts. This system is based on the design of McInteer and Montoya [Recent Developments in Mass Spectrometry in Biochemistry, Medicine and Environmental Research, edited by A. Frigerio (Elsevier, Amsterdam, 1981), Vol. 7, p. 343]. An x-y table and a set of valves under computer control are used to automatically perform hypobromite oxidations, introduce the N2 thus liberated to the mass spectrometer for isotope-ratio analysis, and effect spent sample pumpaway. Analyses can be performed on samples containing 20 μg of N. When operated with a Nuclide Model 3-60-RMS double-collector mass spectrometer, the standard deviation for measurements at the natural abundance level (12 samples, 50 or 100 μg of N per sample) ranged from 0.0002 to 0.000 06 at. % 15N. Throughput capacity is 100–250 samples per day, and unattended operation can continue for at least three days.
8
A Mass Spectrometer for Isotope and Gas Analysis
A. O. Nier · Review of Scientific Instruments · 1947 · 559 citations
D. Rittenberg, Albert S. Keston, F. Rosebury et al. · Journal of Biological Chemistry · 1939 · 98 citations · Full text