Physics in Medicine and Biology · 2002 · 74 citations · 16 references
MedicineBiophysicsRadiologyRadiation DetectionRadiation DosimetryRadiation MeasurementAbsorbed DoseDosimetry InstrumentationIn Vivo DosimetryPhoton DosimetryRadiation Dose MeasurementAttenuation Coefficient MeasurementDiffuse Attenuation CoefficientDosimetersHigh-energy Photon Beam DosimetryHigh-energy Photon DosimetryPolymer Radiation AttenuationVolumetric FlasksX-ray Computed Tomography
Post-irradiation changes in the linear attenuation coefficient, mu, of polymer gel dosimeters give rise to a change which can be measured with x-ray computed tomography. Polymer gel dosimeters were manufactured consisting of 3% (by weight) acrylamide and 3% N,N'-methylene-bis-acrylamide comonomers dissolved in aqueous gelatin (5% gelatin by total weight and 89% de-ionized distilled water). Mu was measured in a collimated radiation beam of photons from an 241Am source. Density, rho, of polymer gel dosimeters was measured using volumetric flasks with capillary stoppers. The measured post-irradiation data of mu was plotted against the data of rho for different batches, and linear least squares fits gave r2 values of 0.99605 and 0.99953, with P values of less than 0.001. This confirms that the post-irradiation change in mu is proportional to that of rho. The change in rho implies a change in volume regardless of the evaluation modality of the polymer gel dosimeter.