JOURNAL OF CHEMICAL ENGINEERING OF JAPAN · 1999 · 17 citations · 4 references
EngineeringPhysiologyComputational Fluid DynamicsFluid DynamicsTransport PhenomenaBiofluid MechanicsCirculatory PhysiologyHemodynamicsHemodynamics ModelingOxygen TransferOxygen TransportBlood FlowBlood Flow SimulationBlood PressureOxygen Transfer ModelingOxygen Transfer RatePorous MediaReynolds NumberSimulation Analysis
An outside blood flow membrane oxygenator with an optimum blood flow pattern is developed by a trial and error method where large cost and long time are spent. We modeled the blood flow in the membrane oxygenator as the flow of low Reynolds number in a porous media and evaluated blood pressure drop and oxygen transfer rate using Darcy’s law. Governing equations are solved by the finite element method. Theoretical blood flow patterns agree well with visualized flow patterns obtained by X-ray computed tomography. Calculated values for blood pressure drop and oxygen transfer rate also agree with experimental data. Optimum structure parameters of the membrane oxygenator can hence be determined by simulation analysis.
4
G. S. Adair, A. V. Bock, H. Field · Journal of Biological Chemistry · 1925 · 823 citations · Full text
Compact cross-flow tubular oxygenators.
Mockros Lf, Ronald Leonard · PubMed · 1985 · 63 citations
Low—order approximations for membrane blood oxygenators
E. N. Lightfoot · AIChE Journal · 1968 · 37 citations