Electronics Letters · 2014 · 15 citations · 4 references
EngineeringBioengineeringBiomedical EngineeringHeat TransferHeatingNeural EngineeringDeep Brain StimulationMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsNeural StimulationNeurostimulationRadio-frequency HeatingDeep Brain Stimulation DevicesDeep Brain Stimulation HardwareLead DesignNeuromodulation Lead DesignRadiofrequency Coil DesignRadiofrequency CoilsRadiofrequency HeatingTemperature Rise Analysis
Radio‐frequency (RF) induced heating at the electrodes of the deep brain stimulation lead is a major concern when used in magnetic resonance imaging (MRI). A novel lead design is proposed to address this problem. By adopting a coil with non‐uniform diameters, the maximal temperature rise measured in a standard phantom is reduced from 4.6 to 2.0°C, with an average reduction of 59%. The suppression of RF heating is probably achieved because of the alteration of the transmission line parameters of the lead by the novel design. The details of the design, experiment validation and discussion of the mechanisms are presented.
4
Ali R. Rezai, Daniel A. Finelli, J.A. Nyenhuis et al. · Journal of Magnetic Resonance Imaging · 2002 · 291 citations · Full text