On the implementation of a recently proposed dosimetric formalism to a robotic radiosurgery system

E. Pantelis, A. Moutsatsos, K. Zourari, W Kilby, Christos Antypas, Panagiotis Papagiannis, P. Karaiskos, Evangelos Georgiou, L. Sakelliou

Medical Physics · 2010 · 61 citations · 32 references

DOIFull text

Open access

Abstract

The comparison of Farmer chamber measurements versus alanine reference dosimetry validates the use of the former for dosimetry in the msr field of this treatment delivery system. The corresponding results of this work obtained using chambers with different cavity lengths, combined with previous literature findings, suggest that a k(Q(msr),Q)(f(msr),f(ref)) Farmer chamber dose response correction factor of 1.01 may improve calibration measurement accuracy when using the proposed dosimetric formalism. The k(Q(msr),Q)(f(msr),f(ref)) correction factor is within 0.5% from unity for ion chambers with cavity lengths less than 10 mm. Substantial improvements in small field output factor measurement accuracy can be obtained when using microchambers and diodes by applying appropriately calculated correction factors to the detector measurements according to the proposed dosimetric formalism, and their routine use is therefore recommended.