Medical Physics · 2011 · 37 citations · 6 references
MedicineOncologyRadiation OncologyClinical Radiation OncologyRadiation DosimetryRadiation MeasurementExternal Beam DosimetryIn Vivo DosimetryProton DosimetryProton TherapyRadiation Dose MeasurementRadiation TherapyCalibration Curve AnalysisCalibration Curve FittingExternal Beam Radiotherapy DosimetryMetal-oxide-semiconductor Field-effect Transistor DosimetryProton Beam Dosimetry CalibrationProton Therapy DosimetrySubmillimeter AccuracyWater-equivalent Depth Measurement
PURPOSE: To verify water equivalent path length (WEPL) before treatment in proton radiotherapy using time resolved in vivo diode dosimetry. METHODS: Using a passively scattered range modulated proton beam, the output of a diode driving a fast current-to-voltage amplifier is recorded at a number of depths in a water tank. At each depth, a burst of overlapping single proton pulses is observed. The rms duration of the burst is computed and the resulting data set is fitted with a cubic polynomial. RESULTS: When the diode is subsequently set to an arbitrary depth and the polynomial is used as a calibration curve, the "unknown" depth is determined within 0.3 mm rms. CONCLUSIONS: A diode or a diode array, placed (for instance) in the rectum in conjunction with a rectal balloon, can potentially determine the WEPL at that point, just prior to treatment, with submillimeter accuracy, allowing the beam energy to be adjusted. The associated unwanted dose is about 0.2% of a typical single fraction treatment dose.
6
<i>Numerical Recipes, The Art of Scientific Computing</i>
William H. Press, Brian P. Flannery, Saul A. Teukolsky et al. · American Journal of Physics · 1987 · 10.1K citations
Numerical Recipes: The Art of Scientific Computing
Eric R. Ziegel · Technometrics · 1987 · 2.3K citations
The art of electronics. 2nd edn