First Laser Endoscopy via a Fiberoptic Transmission System

Gopal Nath, W. Gorisch, P. Kiefhaber

Endoscopy · 1973 · 106 citations · 0 references

Concepts

Abstract

For the first time it was possible to transmit a powerful laser beam (6 watt argon laser) through the biopsy channel of a conventional endoscope. This was achieved by the development of a special quartz fiberoptic transmission system, consisting of a single very flexible fiber with a diameter of 150 µ. A strong absorption effect on liver tissue was observed with the argon laser, due to selective colour absorption of hemoglobin in the green-blue spectral region. The effect on the intact liver surface was studied using different power levels and exposure times. The size of the burned areas and the depth of affected tissue can be exactly controlled by variation of both power and exposure time. After performance of a liver biopsy with the Menghini needle, arrest of hemorrhage was successful in cases with slight and more intense bleedings. Furthermore, the diameter of a diaphragmatical vein shrinked markedly after exposure to a low power beam without getting leakage.