Concepedia

Publication | Closed Access

Clinical and statistical evaluation of self-monitoring blood glucose meters.

69

Citations

3

References

1998

Year

Abstract

Both the Spearman's and Wilcoxon's paired tests, although commonly used, are inappropriate to evaluate SMBG systems. The percentage of SMBG values within the +/-10% interval and the error grid analysis are more accurate, because they consistently classified the five glucose meters tested in our study with a high degree of reproducibility.