Publication | Closed Access
Clinical and statistical evaluation of self-monitoring blood glucose meters.
69
Citations
3
References
1998
Year
Both the Spearman's and Wilcoxon's paired tests, although commonly used, are inappropriate to evaluate SMBG systems. The percentage of SMBG values within the +/-10% interval and the error grid analysis are more accurate, because they consistently classified the five glucose meters tested in our study with a high degree of reproducibility.
| Year | Citations | |
|---|---|---|
Page 1