Design, development, and implementation of the Radiological Physics Center's pelvis and thorax anthropomorphic quality assurance phantoms

D Followill, DeeAnn Radford Evans, Christopher Cherry, A Molineu, Gary D. Fisher, W. F. Hanson, Geoffrey S. Ibbott

Medical Physics · 2007 · 118 citations · 22 references

TL;DR

The study describes the design and development of two anthropomorphic QA phantoms for pelvis IMRT and thorax SBRT and demonstrates their reproducibility. The phantoms are lightweight, water‑fillable, heterogeneous models that mimic patient anatomy, include imageable targets and organs at risk of similar densities, and contain dosimetry inserts for film and TLD, and were imaged, planned, and irradiated at least three times during commissioning. Under controlled irradiation, dose delivery reproducibility was 3 % for the pelvis phantom and 0.5 % for the thorax phantom, field‑localization reproducibility was lower, and the phantoms enable high‑precision verification and credentialing for NCI‑sponsored trials.

Abstract

The Radiological Physics Center (RPC) developed two heterogeneous anthropomorphic quality assurance phantoms for use in verifying the accuracy of radiation delivery: one for intensity‐modulated radiation therapy (IMRT) to the pelvis and the other for stereotactic body radiation therapy (SBRT) to the thorax. The purpose of this study was to describe the design and development of these two phantoms and to demonstrate the reproducibility of measurements generated with them. The phantoms were built to simulate actual patient anatomy. They are lightweight and water‐fillable, and they contain imageable targets and organs at risk of radiation exposure that are of similar densities to their human counterparts. Dosimetry inserts accommodate radiochromic film for relative dosimetry and thermoluminesent dosimetry capsules for absolute dosimetry. As a part of the commissioning process, each phantom was imaged, treatment plans were developed, and radiation was delivered at least three times. Under these controlled irradiation conditions, the reproducibility of dose delivery to the target TLD in the pelvis and thorax phantoms was 3% and 0.5%, respectively. The reproducibility of radiation‐field localization was less than for both phantoms. Using these anthropomorphic phantoms, pelvic IMRT and thoracic SBRT radiation treatments can be verified with a high level of precision. These phantoms can be used to effectively credential institutions for participation in specific NCI‐sponsored clinical trials

References

22