Multiplex TechnologyEngineeringPathogen DetectionBiomedical EngineeringBiosensorsBiosensing SystemsMultichannel WaveguideNanosensorMolecular DiagnosticsMicrofluidicsPlanar Waveguide SensorBiomedical AnalysisBiophotonicsSingle-molecule DetectionBiomolecular EngineeringWaveguide-based Biosensor PlatformBiomedical DiagnosticsMultichannel WaveguidesLab-on-a-chipMicrobiologyQuantitative Multiplex DetectionBiomedical Applications
No single biomarker can accurately predict disease. An ideal biodetection technology should be capable of the quantitative, reproducible, and sensitive detection of a limited suite of such molecules. To this end, we have developed a multiplex biomarker assay for protective antigen and lethal factor of the Bacillus anthracis lethal toxin using semiconductor quantum dots as the fluorescence reporters on our waveguide-based biosensor platform. The platform is extendable to a wide array of biomarkers, facilitating rapid, quantitative, sensitive, and multiplex detection, better than achievable by conventional immunoassay. Our assay allows for the sensitive (limit of detection 1 pM each), specific (minimal nonspecific binding), and rapid (15 min) detection of these biomarkers in complex biological samples (e.g., serum). To address the issue of reproducibility in measurement and to increase our sample throughput, we have incorporated multichannel waveguides capable of simultaneous multiplex detection of biomarkers in three samples in quadruplicate. In this paper, we present the design, fabrication, and development of multichannel waveguides for the simultaneous detection of lethal factor and protective antigen in serum. Evaluation of the multichannel waveguide shows an excellent concordance with single-channel data and effective, simultaneous, and reproducible measurement of lethal toxins in three samples.
11
Multiplexed Toxin Analysis Using Four Colors of Quantum Dot Fluororeagents
Ellen R. Goldman, Aaron R. Clapp, George P. Anderson et al. · Analytical Chemistry · 2003 · 648 citations
Patrick M. Winter, Shelton D. Caruthers, Andrea Kassner et al. · PubMed · 2003 · 376 citations
Designing a polyvalent inhibitor of anthrax toxin
Michaël Mourez, Ravi S. Kane, Jeremy Mogridge et al. · Nature Biotechnology · 2001 · 270 citations
Waveguide-Based Biosensors for Pathogen Detection
Harshini Mukundan, Aaron S. Anderson, W. Kevin Grace et al. · Sensors · 2009 · 190 citations · Full text