The HeartMate II: Design and Development of a Fully Sealed Axial Flow Left Ventricular Assist System

David Burke, E.A. Burke, Farzad Parsaie, Victor L. Poirier, K. Butler, Douglas J. Thomas, L. Taylor, Timothy Maher

Artificial Organs · 2001 · 76 citations · 4 references

DOIFull text

Open access

Concepts

Abstract

Our group is developing the control and power transmission components required to implement a permanent and fully sealed left ventricular assist system (LVAS). Starting with the percutaneously powered HeartMate II blood pump, our development efforts are focused in the following areas: a complete redesign of the transcutaneous energy transmission system (TETS) to include a rectification network and autonomous voltage regulation within the secondary coil, a hermetically sealed electronics package containing a miniaturized implementation of the existing redundant drive and control electronics with several power-input options, an implanted rechargeable lithium ion battery pack capable of providing up to 1 h of untethered operation, implantable electrical connectors that allow components to be connected after placement in the body or to be replaced if needed, and a radio telemetry subsystem to transmit diagnostic information and to permit remote adjustment of selected parameters.

References

4