Review of Scientific Instruments · 1965 · 24 citations · 9 references
EngineeringNuclear EngineeringPhysicsNuclear PhysicsRadiation DetectionRadiation PhysicsDetector TechnologyGamma-ray DetectionGamma-ray DetectorsNeutron DetectionRadiation DetectorsScintillation DetectorsLiquid ScintillatorsNeutron Detector SystemsNeutron DetectorsPulse CharacteristicsPulse DiscriminationPulse Shape AnalysisPulse-shape DiscriminationScintillation Neutron Detectors
A boron-loaded liquid scintillator is described which gives a much greater difference in the pulse shapes for γ rays and for slow neutrons than any of the liquids currently used. A detailed study of the pulse shapes indicates that the γ-ray sensitivity of a neutron detector filled with this liquid can be reduced by a factor of 500 with only a 5% decrease in the neutron sensitivity by application of pulse-shape discrimination. A survey of the pulse-shape characteristics of a number of scintillators currently used, together with the results for the new liquid, is presented, and the performance of the new detector with a simple system of pulse-shape discrimination is described.
9
A New Quenching Effect in Liquid Scintillators
Robert Pringle, L. D. Black, B. Lionel Funt et al. · Physical Review · 1953 · 56 citations