Publication | Closed Access
A Nonlocking End Screw Can Decrease Fracture Risk Caused by Locked Plating in the Osteoporotic Diaphysis
124
Citations
24
References
2009
Year
Under bending loads, the focused load transfer of locking plates through fixed-angle screws can increase the periprosthetic fracture risk in the osteoporotic diaphysis compared with conventional plates. Replacing the outermost locking screw with a conventional screw reduced the stress concentration at the plate end and significantly increased the bending strength of the plating construct compared with an all-locked construct (p = 0.001).
| Year | Citations | |
|---|---|---|
Page 1