Journal of Chromatographic Science · 1980 · 41 citations · 0 references
PharmacyChemistryPharmaceutical AnalysisAnalytical ChemistryChromatographic AnalysisMass Spectrometric AnalysisAnalytical ChromatographyBioanalytical Method ValidationChromatographic Data AnalysisChromatographyLiquid ChromatographyPeak Concentration AnalysisAnalytical Liquid ChromatographyElectrode Potential MeasurementFlow Rate DependenceHigh-performance Liquid ChromatographyLower Limit Of QuantificationLower Limit Of QuantitationNitrosamine Impurity Analysis
The polarographic detector for high performance liquid chromatography has been evaluated with respect to a model compound, N-nitrosodipropylamine (DPN), and composite mixtures of volatile and nonvolatile nitrosamines. The effect of electrode potential, drop time, drop size, pH, and flow rate has been investigated in the differential pulse and sampled DC modes. Plots of amount injected vs. peak height are linear over three-four decades of concentration with a lower limit of 0.8 ng.