A new method to study the efficiency of third generation blood filters

M. Sivakumaran, D. R. Norfolk, Kimberly Major, J. A. Revill, R. M. Hutchinson, J. K. Wood

British Journal of Haematology · 1993 · 14 citations · 2 references

Concepts

Abstract

In order to study the efficiency of third generation blood filters we have devised a new method using 3 microns pore size polycarbonate filter membranes and a filtration chamber to trap leucocytes passing through the blood filter. The method has enabled us to make two important observations: (1) The filters function very efficiently at the start but the efficiency diminishes progressively with time. (2) When the blood flow through the filter is retarded, due to defective priming and/or filter malfunction, the filters fail even at the outset of the transfusion.