2012 · 20 citations · 4 references
EngineeringBioengineeringBiomedical EngineeringCochlear ImplantsElectrophysiologyNeural EngineeringAuditory Nerve StimulationCochlear ImplantationImplant TechnologyImplantable Biomedical DevicesNeurostimulationSurgical ImplantationCochlear Electrode-cochlea InterfaceCochlear ElectrodesCochlear Implant StimulationFeedback TopologyLoop TopologyNegative Feedback TopologiesOutput ImpedanceOutput Impedance Control
This paper presents an implementation of a least voltage drop neural biphasic stimulator circuit applied in cochlear implants. Using a double loop negative feedback topology, the output impedance of the current generator is increased, while requiring only one effective drain-source voltage drop (V <inf xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">eff</inf> ). This allows the circuit to convey more charge into the tissue. The circuit can provide a biphasic stimulation scheme from a single ended supply with an amplitude range of 10µA up to 1.05mA for a wide range of electrode-tissue impedances, R <inf xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">L</inf> =1kΩ∼10kΩ, C <inf xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">L</inf> =1nF∼10nF. The stimulation current is set by scaling a reference current using a two stage binary-weighted transistor DAC configuration (3 bits HV transistor DAC and 4 bits LV transistor DAC) to improve the speed of stimulation pulses and minimize the area of the circuit. Simulation results, using the AMS 0.18µm high-voltage CMOS process, show that the charge error within a cycle (600µs) is only 0.02%, equivalent to a DC current error of 3nA at the maximum stimulation current with a load of 10kΩ+10nF.
4
Ji‐Jon Sit, Rahul Sarpeshkar · IEEE Transactions on Biomedical Circuits and Systems · 2007 · 153 citations
Engineering, Biomedical Engineering, Electrical Engineering +17