Engineering in Medicine · 1982 · 23 citations · 7 references
Natural SciencesMaterials EngineeringOrthopedic SurgeryCorrosion ScienceMetallurgyOrthopedic BiomaterialsCorrosion ResistanceCorrosion TestingCorrosion-resistant MaterialsJoint Replacement MaterialsCorrosion FatigueCorrosion Fatigue TestingCorrosion-fatigue InteractionFatigue StrengthOrthopaedic ImplantsPowder Metallurgy AlloysShot PeeningSteam Sterilization
The paper compares four alloys in orthopaedic implant use on the basis of their performance in laboratory tests. The alloys compared were cast Co–Cr–Mo, powder metallurgy (PM) Co–Cr–Mo, Ti 318 and Ti 550. Tensile, corrosion fatigue, and corrosion tests were carried out and the effect of heat treatment and of surface treatment was investigated. The corrosion fatigue tests ranked the alloys in the following order of merit (1) Co–Cr–Mo (PM), (2) Ti 550, (3) Ti 318, and (4) cast Co–Cr–Mo. The corrosion tests demonstrated for cast Co–Cr–Mo the effect of passivation, steam sterilization, and texturing, and for Ti 318 the effect of oxidizing, anodizing, and acid etch/shot peening. It was concluded that for joint replacement applications requiring the highest corrosion fatigue strength regardless of cost and neglecting the possible importance of elastic modulus, Co–Cr–Mo (PM) was best. For applications requiring ease of manufacture and allowing a lower corrosion fatigue strength, cast Co–Cr–Mo was the material of choice. For most other applications and in particular where high corrosion fatigue strength at a reasonable cost was required, the Ti alloys offered worthwhile advantages.
7
Survivorship of total hip replacements
H. S. Dobbs · Journal of Bone and Joint Surgery - British Volume · 1980 · 289 citations
Metal ion release after total hip replacement
H. S. Dobbs, M.J. Minski · Biomaterials · 1980 · 150 citations
Fracture of titanium orthopaedic implants
H. S. Dobbs · Journal of Materials Science · 1982 · 31 citations