Natural SciencesMaterials EngineeringMaterials ScienceNuclear PhysicsRadiation MeasurementThin FilmsFilm DosimetryMaterial FilmNeutron DetectionThin Film Thermal PropertiesFast Neutron DetectionFast Neutron SpectraNeutron Radiation AssessmentPhotographic Film DosimetryPrompt NeutronsThermal Neutron DetectionThermal Neutron SpectraThermal Neutron Spectrum
Measurements were made to determine the sensitivity to 3-MeV neutrons and to thermal neutrons of three commercial photographic films (DuPont film types 502, 510 and 606), The films were exposed in the conventional “NBS holders”, consisting of Bakelite covered with thin layers of lead and tin. The results of the measurements were compared with results obtained by rough theoretical estimates. The sensitivity of the films to neutrons was found to be relatively low. On the least sensitive film, a photographic density of 0.1 was obtained with a fast-neutron exposure of 6.8 × 1010 n/cm2. On the most sensitive film, the exposure for the same density was 4.9 × 109 n/cm2. The ratios of the films' fast-neutron-to-gamma-ray sensitivities were found to be roughly- equal to the reciprocal of the ratios of their average grain diameters. The thermal neutron sensitivity of the films was of the same order as the 3-MeV sensitivity, and was roughly the same for all three film types. In the few cases in which values for the neutron sensitivity could be obtained theoretically, they showed order-of magnitude agreement with the experimental values.