Influence of the Test Area on the Mechanical Properties of Skin
98
Citations
0
References
1995
Year
The Cutometer is a reliable device. A high degree of correlation exists between biomechanical variables related to elasticity. Under a suction of 500 mbar, both the 2- and 8-mm probes give results which correlate statistically in a large population. However, the two probes do not measure precisely the same aspect of skin mechanics. The 2-mm probe measures the capacity of superficial skin folding, and the 8-mm probe explores the biomechanical properties of the dermis itself and of its slipping mobility on the hypodermis. The biomechanical parameters which appear the most clinically relevant for a normal-looking skin are the maximum distension and the biological elasticity.