Single- and dual-energy CT with monochromatic synchrotron x-rays

110

Citations

21

References

1997

Year

Abstract

We explored the potential for clinical research of computed tomography (CT) with monochromatic x-rays using the preclinical multiple energy computed tomography (MECT) system at the National Synchrotron Light Source. MECT has a fixed, horizontal fan beam with a subject apparatus rotating about a vertical axis; it will be used for imaging the human head and neck. Two -photodiode array detectors with different spatial resolutions were used. A 10.5 cm diameter acrylic phantom was imaged with MECT at 43 keV and with a conventional CT (CCT) at 80 kVp: spatial resolution line pairs (lp)/cm for both; slice height, 2.6 mm for MECT against 3.0 mm for CCT; surface dose, 3.1 cGy for MECT against 2.0 cGy for CCT. The resultant image noise was 1.5 HU for MECT against 3 HU for CCT. Computer simulations of the same images with more precisely matched spatial resolution, slice height and dose indicated an image - noise ratio of 1.4:1.0 for CCT against MECT. A 13.5 cm diameter acrylic phantom imaged with MECT at above the iodine K edge and with CCT showed, for a solution, an image contrast of 26 HU for MECT and 13 and 9 HU for the 80 and 100 kVp CCT, respectively. The corresponding numbers from computer simulation of the same images were 26, 12, and 9 HU, respectively. MECT's potential for use in clinical research is discussed.