British Journal of Radiology · 1995 · 17 citations · 4 references
Computed TomographyTechnique OptimizationMedical ImagingLow DoseCraniofacial DisorderPediatricsDiagnostic NeuroradiologyPaediatric RadiologyCt ScanSurgeryLens DoseAnatomyCraniofacial SurgeryMedicineOrthopaedic SurgeryRadiologyHealth Sciences
21 infants with craniosynostosis were studied with a new three-dimensional (3D) computed tomography (CT) methodology. We describe technique optimization using multiple spiral data acquisitions with low dose (85 mAs) technique. One caudal volume of 3 mm slice thickness was obtained with a further two volumes of 1 mm slice thickness at the vertex. Image reconstruction of spiral raw data allowed overlapping 3 mm sections to be generated without the dose increase that would result from conventional axial CT scanning. We illustrate common technical artefacts of 3D CT and explain their cause and solution. A dramatic dose reduction to the lens was achieved with no loss in 3D image quality. Lens dose was 8.91 mSv compared with 24.6 mSv using the standard paediatric head technique.
4
Value of 3D imaging for the study of craniofacial malformations in children.
Nicolas Leboucq, Pedro García-Alonso Montoya, Yannick Martinez et al. · PubMed · 1991 · 14 citations