Optics Letters · 1993 · 134 citations · 5 references
A white-light interferometric extrinsic Fabry-Perot strain sensor that uses a multimode fiber is demonstrated. The Fabry-Perot cavity length is measured with the help of a Fizeau interferometer. The sensor is described, and some results obtained at this time are given. The strain measurements are absolute and perfectly linear, with a sensitivity of 0.25 micrometers per meter (micro). the design of a thermally autocompensated strain sensor is also presented.
5
Quadrature phase-shifted, extrinsic Fabry–Perot optical fiber sensors
Kent A. Murphy, Michael F. Gunther, Ashish M. Vengsarkar et al. · Optics Letters · 1991 · 418 citations
Coherence multiplexing of remote fibre optic fabry-perot sensing system