Concepedia

Publication | Closed Access

Comparison of DVH data from multiple radiotherapy treatment planning systems

68

Citations

16

References

2010

Year

Abstract

This study examined the variation of dose-volume histogram (DVH) data sourced from multiple radiotherapy treatment planning systems (TPSs). Treatment plan exports were obtained from 33 Australian and New Zealand centres during a dosimetry study. Plan information, including DVH data, was exported from the TPS at each centre and reviewed in a digital review system (SWAN). The review system was then used to produce an independent calculation of DVH information for each delineated structure. The relationships between DVHs extracted from each TPS and independently calculated were examined, particularly in terms of the influence of CT scan slice and pixel widths, the resolution of dose calculation grids and the TPS manufacturer. Calculation of total volume and DVH data was consistent between SWAN and each TPS, with the small discrepancies found tending to increase with decreasing structure size. This was significantly influenced by the TPS model used to derive the data. For target structures covered with relatively uniform dose distributions, there was a significant difference between the minimum dose in each TPS-exported DVH and that calculated independently.